- Provides 700VA, 450watt(s) power / Surge Suppression is Standard
- 18 minute(s) Battery Runtime at Full Load
Tripp Lite's SMART700HG is UL2601-1 listed as medical electrical equipment with less than 300 microamps leakage current for the UPS and all connected equipment. Built-in isolation transformer with faraday shield removes common mode noise while maintaining full AC surge suppression in all modes. It is also resistant to electrostatic discharge. This UPS includes high quality hospital-grade plug and outlets. 700VA power handling capacity supports computers, internetworking equipment and other critical loads during commercial power failures. Line interactive voltage regulation corrects brownouts as low as 81V and overvoltages as high as 143 back to normal 120V levels, saving battery reserves for use only during complete power failures. This model includes 4 outlets, diagnostic LEDs and built-in USB and RS-232 communications ports that work with downloadable PowerAlert Software and cabling to provide automatic unattended shutdown of connected network and PC applications.
|
General
|
|
|
Form Factor
|
External
|
|
Power
|
|
|
Surge Suppression
|
Standard
|
|
Power Provided
|
700 VA
|
|
Power Provided
|
450 watt(s)
|
|
Battery Run Time (max)
|
18 minute(s)
|
|
Actual Weight
|
28.00 lb(s)
|
|
Compatibility Notes
|
UL 2601-1 listed UPS provides ultimate safety and reliability for patient point of care solutions, telemedicine, nurse's stations, wireless communications systems, electronic prescribing applications, medical imaging and mission critical equipment. Not recommended for life support applications.
Voltage compatibility: 120V AC Frequency compatibility: 60 Hz
|
|
Contents
|
UPS tower, cables and documentation
|
|
Physical Dimensions
|
9"h x 12.5"w x 7.2"d
|
|
Ports/Connectors
|
(4) NEMA 5-15R Hospital Grade outlets (1) NEMA 5-15P Hospital Grade Input cord: 6 feet 14 gauge (1) SNMP expansion slot (1) USB port (1) RS-232 port
|
|
Power Notes
|
OUTPUT
Output volt amp capacity (VA): 700 Output watt capacity (watts): 450 Output nominal voltage: 120V
Output voltage regulation:
LINE MODE: Sine wave line voltage 120V (-18 percent / +8 percent) BATTERY MODE: PWM sine wave output 120V +/-5 percent
Output frequency regulation:
LINE MODE: Passes line frequency of 60Hz +/-10 percent BATTERY MODE: Inverter output regulated to 60Hz +/-0.5Hz
Overload protection: Resettable circuit breaker
INPUT
Maximum input amps: 5.5A / 550 watts Recommended electrical service: 15A 120V
BATTERY Full load runtime: 18 minutes (700VA) Half load runtime: 42 minutes (350VA) DC system voltage: 36V DC Typical battery lifespan: 3-6 years, depending on usage Battery recharge rate: 2-4 hours to 90 percent
VOLTAGE REGULATION
Voltage regulation description: Line interactive UPS offers 2 boost levels of brownout correction and a single level of overvoltage adjustment to maintain usable, computer grade output without consuming battery power over an input voltage range of 81 to 143V AC.
Overvoltage correction: Input voltages between 126 and 143V AC are reduced by 9 percent Direct pass through: Input voltages between 105 and 125V AC are passed through unchanged
Brownout correction: Input voltages between 96 and 106V AC are boosted by 10 percent
Severe brownout correction: Input voltages between 81 and 95V AC are boosted
SURGE / NOISE SUPPRESSION
AC surge suppression: Conforms to IEEE 587 / ANSI C62.41 specifications AC suppression response time: Instantaneous EMI / RFI AC noise suppression: Yes
|
|
Returns Policy
|
This product is subject to our return policy. Please see our complete return policy for details.
|
|
Warranty - Labor
|
2 Years
|
|
Warranty - Parts
|
2 Years
|
|
Warranty Details
|
2-year warranty, 3 year with registration. Note: registration is required for 3-year warranty. Connected Equipment Insurance (U.S., Canada & Puerto Rico) $250, 000 Ultimate Lifetime Insurance
|
http://www.macconnection.com/IPA/Shop/Product/Detail.htm?sku=133716#BVSecondaryCustomerRatings